PRODUCT NAME
Gasboard-3200 Plus

Specification

Gasboard-3200 Plus
Measurement CH4, CO2 (NDIR), O2, H2S(ECD)
Option CO, H2 (ECD), gas flow(Thermal mass flow probe)
Range CH4: (0~100)%, CO2: (0~50)%, O2: (0~25)%, H2S: (0~9999) ppm, CO: (0~2000) ppm, H2: (0~2000) ppm
Precision CH4, CO2: ≤2%FS, O2, H2S, CO, H2: ≤3%FS
Resolution CH4, CO2, O2: 0.01%, H2S, CO, H2: 1ppm
GPS sensor Positioning and location
Lithium battery pack Rechargeable
Display High-resolution colored 3.2-inch
Inlet pressure 2KPa~50KPa
Communication USB port, blue tooth 4.0
Dimension 276 x 195 x 66 mm

Infrared On-Line Biogas Analyzer

RODUCT NAME
Gasboard-3200

Specification

Gasboard-3200
Measurement CH4, CO2 (NDIR), O2 , H2S (ECD)
Range CH4 : (0 - 100)%, CO2 : (0 - 50)%, O2 : (0 - 25)%, H2S : (0 - 9999)ppm
Precision CH4, CO2: ≤2%FS, O2, H2S, CO, H2: ≤3%FS
Resolution CH4, CO2, O2 : 0.01%, H2S : 1ppm
Repeatability ≤2%
Inlet pressure 2KPa~50KPa
Sampling gas condition No dust, no vapor
Response time(T) 10s (NDIR)
Communication RS232
Output (4~20)mA
Power supply 110V/220V±44V AC 60Hz±1Hz
Dimension 485mm x 457mm x 132mm
Weight Approx.12kgs
Infrared Biogas Analysis System Gasboard-9060 Infrared Biogas Analysis System Gasboard-9060 is designed for the measurement of the composition of CH4 CO2 H2S and O2 in the landfill gas and biogas. It is based on the modern NDIR and ECD technology. The gas sampling from the normal pressure or mild under pressure is made by a membrane pump,from the overpressure by a needle valve. The gas treatment unit is offered to remove humidity and dust particles from the gas, as well as long lifespan H2S protection module to ensure the safety of the system reliable operation.

Infrared BiogasAnalysis System

RODUCT NAME
Gasboard -9060

Specification

Gasboard-9060
Measurement range CH4: (0~100)%, CO2: (0~50)%, H2S: (0~9999)ppm, O2: (0~25)%
Precision CH4, CO2: ≤2%FS, H2S, O2: ≤3%FS
Resolutions 0.01% for CO2, CH4, O2 and 1ppm for H2S
Characteristic

Auto zeroing function

Pressure compensation function

With long life time H2S protection device

The stainless steel cabinet and automatic air purge on H2S sensor

The overall system with temperature control, suitable for low temperature